Logo Data Input

Nutribox—The New Generation

Impedance Analysis Apparatus Nutribox Technical Data

RESISTANCE (R)  
Area of measurement 0 - 999 Ohm
Resolution of measurement 1 Ohm
Exactness of measurement ± 1 % of measurement value
± 1 Digit.
REACTANCE (Xc)  
Area of measurement 0 - 250 Ohm
Resolution of measurement 1 Ohm
Exactness of measurement ± 2.5 % of measurement value
± 1 Digit

The exact details given for the measurement values are only valid for opposition in ohm-based, high capacity components.

Measurement Current 0,8 mA bei 50 kHz (0 - 1000 Ohm)
Exactness of the measurement current ± 3 %
Measurement frequency 50 kHz sinusoidal frequency
Source of electrical current via PC or measurement via rechargeable battery
Apparatus type Impedance-Analyser
Apparatus name Nutribox
Class of protection (Nutribox) Apparatus with electrical current via PC&rquo;s USB port
Degree of protection (Nutribox) Type B

CE (Conformité Européene) in accordance with EN55022 Class B